Correlation Engine 2.0
Clear Search sequence regions


Sizes of these terms reflect their relevance to your search.

A method for the determination of ultratrace amounts of Cr, Fe, Mn, Pb and Zn in seawater has been developed. It combined the low-blank magnesium hydroxide coprecipitation procedure with quadrupole inductively coupled plasma mass spectrometry and used the dynamic reaction cell technique to resolve the polyatomic interferences arising from the residual matrix, the solvent and plasma gases. Detection limits (3σ(B), n=10) for Cr, Fe, Mn, Pb and Zn were 0.02, 0.10, 0.01, 0.002 and 0.19 nM, respectively, using 50 mL of seawater sample. The accuracy of the analytical procedure was verified by the analysis of the seawater reference materials CASS-4, NASS-5, SAFe D2 and SAFe S. The analytical precision ranged from 3% to 16% (n=6), with a sample throughput of about 6 samples h(-1). Copyright © 2011 Elsevier B.V. All rights reserved.

Citation

Francisco Ardini, Emanuele Magi, Marco Grotti. Determination of ultratrace levels of dissolved metals in seawater by reaction cell inductively coupled plasma mass spectrometry after ammonia induced magnesium hydroxide coprecipitation. Analytica chimica acta. 2011 Nov 7;706(1):84-8

Expand section icon Mesh Tags

Expand section icon Substances


PMID: 21995913