Renata Cléia Claudino Barbosa, Celso Ricardo Fernandes de Carvalho, Henrique Takachi Moriya
Department of Physical Therapy and Biomedical Engineering Laboratory, University of São Paulo, Avenida Dr. Arnaldo 455, São Paulo, Brazil.
Jornal brasileiro de pneumologia : publicaça̋o oficial da Sociedade Brasileira de Pneumologia e Tisilogia 2012 Mar-AprTo compare two methods of respiratory inductive plethysmography (RIP) calibration in three different positions. We evaluated 28 healthy subjects (18 women and 10 men), with a mean age of 25.4 ± 3.9 years. For all of the subjects, isovolume maneuver calibration (ISOCAL) and qualitative diagnostic calibration (QDC) were used in the orthostatic, sitting, and supine positions. In order to evaluate the concordance between the two calibration methods, we used ANOVA and Bland-Altman plots. The values of the constant of proportionality (K) were significantly different between ISOCAL and QDC in the three positions evaluated: 1.6 ± 0.5 vs. 2.0 ± 1.2, in the supine position, 2.5 ± 0.8 vs. 0.6 ± 0.3 in the sitting position, and 2.0 ± 0.8 vs. 0.6 ± 0.3 in the orthostatic position (p < 0.05 for all). Our results suggest that QDC is an inaccurate method for the calibration of RIP. The K values obtained with ISOCAL reveal that RIP should be calibrated for each position evaluated.
Renata Cléia Claudino Barbosa, Celso Ricardo Fernandes de Carvalho, Henrique Takachi Moriya. Respiratory inductive plethysmography: a comparative study between isovolume maneuver calibration and qualitative diagnostic calibration in healthy volunteers assessed in different positions. Jornal brasileiro de pneumologia : publicaça̋o oficial da Sociedade Brasileira de Pneumologia e Tisilogia. 2012 Mar-Apr;38(2):194-201
PMID: 22576427
View Full Text