Sometimes in-vitro diagnostic devices (e.g. blood collection tubes) are not validated before use or when the producer's brand is changed. The aim of this study was to validate five brands of sodium citrate vacuum tubes. Blood specimens from 50 volunteers were collected in five different tube brands (I: Venosafe, II: VACUETTE, III: BD Vacutainer, IV: LABOR IMPORT and V: S-Monovette). Routine coagulation tests [activated partial thromboplastin time (aPTT), prothrombin time (PT), and fibrinogen (FIB)] were performed on ACL TOP instrument using HemosIL reagents. The significance of the differences between samples was assessed by paired Student's t-test, set at Pā€Š<ā€Š0.005. Significant differences were observed for: PT when comparing I vs. II, I vs. III, I vs. V, II vs. III, II vs. IV, II vs. V, III vs. IV, III vs. V and IV vs. V; aPTT when comparing I vs. II, I vs. III, I vs. IV, II vs. IV, III vs. IV and IV vs. V. No differences were observed among brands for FIB determination. We suggest that every laboratory management should both standardize the procedures and frequently evaluate the quality of in-vitro diagnostic devices.

Citation

Gabriel Lima-Oliveira, Giuseppe Lippi, Gian Luca Salvagno, Martina Montagnana, Geraldo Picheth, Gian Cesare Guidi. Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing. Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis. 2013 Apr;24(3):252-5