Correlation Engine 2.0
Clear Search sequence regions


Sizes of these terms reflect their relevance to your search.

Listeria monocytogenes (LM), an important food-borne pathogen that has a high mortality rate (≈ 30%), was successfully detected within an hour at the infection dose limit of 10(3)/mL, both in buffer and milk. LM detection was demonstrated using a novel asymmetrically anchored cantilever sensor and a commercially available antibody. Sensor responses were confirmed using a secondary antibody binding step, similar to the sandwich ELISA assays, as a means of signal amplification that also reduced the occurrence of false negatives. Detection of LM at a concentrations of 10(2)/mL was achieved, by incorporating a third antibody binding step, which is an order of magnitude smaller than the infection dose (<1000 cells) for LM. The commercially available antibody for LM used in this work is shown to have low avidity which partially explains the relatively low sensitivity reported for LM as compared to other pathogens. Copyright © 2013 Elsevier B.V. All rights reserved.

Citation

Harsh Sharma, Raj Mutharasan. Rapid and sensitive immunodetection of Listeria monocytogenes in milk using a novel piezoelectric cantilever sensor. Biosensors & bioelectronics. 2013 Jul 15;45:158-62

Expand section icon Mesh Tags

Expand section icon Substances


PMID: 23475140

View Full Text