Correlation Engine 2.0
Clear Search sequence regions


  • freedom (2)
  • tendon (2)
  • trocar (1)
  • yaw joint (1)
  • Sizes of these terms reflect their relevance to your search.

    This paper describes a multitasking robotic platform for Minimally Invasive Surgery (MIS). The device is designed to be introduced through a standard trocar port. Once the device is inserted to the desired surgical site, it can be reconfigured by lifting an articulated section, and protruding two tendon driven flexible arms. Each of the arms holds an interchangeable surgical instrument. The articulated section features a 2 Degrees-of-Freedom (DoF) universal joint followed by a single DoF yaw joint. It incorporates an on-board camera and LED light source at the distal end, leaving a Ø3mm channel for an additional instrument. The main shaft of the robot is largely hollow, leaving ample space for the insertion of two tendon driven flexible arms integrated with surgical instruments. The ex-vivo and in-vivo experiments demonstrate the potential clinical value of the device for performing surgical tasks through single incision or natural orifice transluminal procedures.

    Citation

    Jianzhong Shang, Christopher J Payne, James Clark, David P Noonan, Ka-Wai Kwok, Ara Darzi, Guang-Zhong Yang. Design of a Multitasking Robotic Platform with Flexible Arms and Articulated Head for Minimally Invasive Surgery. Proceedings of the ... IEEE/RSJ International Conference on Intelligent Robots and Systems. IEEE/RSJ International Conference on Intelligent Robots and Systems. 2012 Oct 12;2012:1988-1993


    PMID: 24748998

    View Full Text