Correlation Engine 2.0
Clear Search sequence regions


  • doripenem (1)
  • human (3)
  • meropenem (6)
  • methanol (2)
  • patient (1)
  • phase (2)
  • serum (4)
  • serum protein (1)
  • volunteers (1)
  • Sizes of these terms reflect their relevance to your search.

    In this study, a new, sensitive and selective high-performance liquid chromatographic method was developed for the determination of meropenem (MEM) in human serum. In the developed method, C18 column (3.9 × 150 mm, 5 μm) was selected as stationary phase at 30°C, and methanol: acetic acid solution mixture was used as mobile phase with gradient program. Chromatographic separation was carried out at a flow rate of 1 mL/min, and detection was performed at 300 nm with diode array detector. Doripenem was selected as an internal standard, and the analytes were extracted from serum using protein precipitation method with ortho-phosphoric acid: methanol. Detection wavelength was selected as 300 nm. The developed method was validated according to International Council for Harmonisation (ICH) guidelines. The calibration curve was linear over a concentration range of 4-240 μg/mL with correlation coefficient of 0.9985. The limit of detection and limit of quantification values were found as 0.057 and 0.192 μg/mL, respectively. The validated method was successfully applied for the determination of MEM in human serum samples collected from patient volunteers at different time intervals, and therapeutic drug monitoring of MEM has been investigated. © The Author(s) 2019. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

    Citation

    Demet Dincel, Olcay Sagirli, Gulacti Topcu. A High-Performance Liquid Chromatographic Method for the Determination of Meropenem in Serum. Journal of chromatographic science. 2020 Jan 23;58(2):144-150

    Expand section icon Mesh Tags

    Expand section icon Substances


    PMID: 31738410

    View Full Text