Correlation Engine 2.0
Clear Search sequence regions


  • brain (3)
  • HISCL (10)
  • hook (1)
  • humans (1)
  • nt probnp (21)
  • serum (3)
  • Sizes of these terms reflect their relevance to your search.

    The aim of the study was to assess the analytical performance of the HISCL NT-proBNP assay, a newly developed chemiluminescence immunoassay, for the detection of NT-proBNP. The within-run and total imprecision of the NT-proBNP assay were determined with HISCL cardiac marker controls. The linear ranges of the NT-proBNP assays were evaluated based on the CLSI EP6-A document using selected serum samples. Two hundred serum samples were evaluated to compare the HISCL NT-proBNP and Elecsys NT-proBNP assays. Five additional high NT-proBNP concentrations serum samples were evaluated to assess if there was high-dose hook effect in the HISCL NT-proBNP assay. The total and within-run imprecision values of the HISCL NT-proBNP assay were 5.85%, 0.81%, 2.56% and 0.54% and 6.07%, 0.73%, 2.61% and 0.59% at 6.1, 129.83, 3732.84and39737.33 pg/ml, respectively. The assay was verified to be linear for NT-proBNP levels ranging between 6.1 and 39737.33 pg/ml. The assay comparison showed that HISCL NT-proBNP = 0.9803 × Elecsys NT-proBNP -4.383. The sensitivity of HISCL NT-proBNP was 87.23%, and the specificity was 85.61%. The AUC of HISCL NT-proBNP (0.90 (95% CI, 0.86-0.93)) did not differ from that of Elecsys NT-proBNP(0.89 (95% CI, 0.85-0.93)) (P = 0.638). The results of five high NT-proBNP concentrations samples (44448, 54206, 55634, 55728 and 109406 pg/ml, measured with the Elecsys NT-proBNP assay) tested with HISCL NT-proBNP assay were all displayed with ">40000 pg/ml". The HISCL NT-proBNP chemiluminescence immunoassay showed good analytical and diagnostic performance for the detection of NT-proBNP and could be used in routine clinical practice. Copyright © 2020 Elsevier Inc. All rights reserved.

    Citation

    Yu-Yan Chen, Shu-Lian Li, Hui-Ling Lin, Wen-Dong Li, Xiao-Zhen Zhu, Hui-Lin Zhang. A chemiluminescence immunoassay for the detection of NT-proBNP. Analytical biochemistry. 2020 Dec 15;611:113950

    Expand section icon Mesh Tags

    Expand section icon Substances


    PMID: 32919941

    View Full Text