Adverse reactions to metal debris are significant complications after metal-on-metal total hip arthroplasty. Recently, late appearances of adverse reactions to metal debris and subsequent need for reoperations have been reported with small-diameter head metal-on-metal devices. We retrospectively investigated mid-term clinical outcomes of small-head metal-on-metal total hip arthroplasty. We reviewed 159 hips in 139 patients who had a small-head metal-on-metal total hip arthroplasty (M2a Taper; Biomet, Warsaw, IN) with a minimum 5-year follow-up and documented postoperative complications. Focal osteolysis in either the femur or acetabulum was observed in 12 hips (7.5%, 44 months after surgery on average), with pseudotumor observed in 8 hips (5%, 120 months after surgery on average). Four hips (2.5%) had dislocations (84 months after surgery on average) and six hips (3.8%, 122 months after surgery on average) underwent reoperation. Small-head metal-on-metal total hip arthroplasty is associated with a high degree of complications at mid-term follow-up period. Considering this, we discourage the use of metal-on-metal total hip arthroplasty regardless of head size. © The Author(s) 2021.

Citation

Nobuhiko Sumiyoshi, Kazuhiro Oinuma, Yoko Miura. Small-head metal on metal total hip arthroplasty is associated with a high rate of complication and reoperation at mid-term follow-up. SAGE open medicine. 2021;9:20503121211014707


PMID: 34035911