Correlation Engine 2.0
Clear Search sequence regions


Sizes of these terms reflect their relevance to your search.

Considering the limitations of well-known traditional detection techniques, innovative research studies have focused on the development of new sensors to offer label-free, highly sensitive, real-time, low-cost, and rapid detection for biomolecular interactions. In this study, we demonstrate immunoglobulin G (IgG) detection in aqueous solutions by using real-time and label-free kinetic measurements of the Interferometric Reflectance Imaging Sensor (IRIS) system. By performing kinetic characterization experiments, the sensor's performance is comprehensively evaluated and a high correlation coefficient value (>0.94) is obtained in the IgG concentration range of 1-50 μg/mL with a low detection limit (0.25 μg/mL or 1.67 nM). Moreover, the highly sensitive imaging system ensures accurate quantification and reliable validation of recorded binding events, offering new perspectives in terms of direct biomarker detection for clinical applications. Copyright © 2022 Elsevier B.V. All rights reserved.

Citation

Monireh Bakhshpour, Elisa Chiodi, Iris Celebi, Yeşeren Saylan, Nese Lortlar Ünlü, M Selim Ünlü, Adil Denizli. Sensitive and real-time detection of IgG using interferometric reflecting imaging sensor system. Biosensors & bioelectronics. 2022 Apr 01;201:113961

Expand section icon Mesh Tags

Expand section icon Substances


PMID: 35026547